407355 Version 2.4 9/07 12
EVENT WINDOW - Download data from meter bank (E1 through E5) to PC
In this window the user can download, view, and save the contents of the meter’s Event
banks E1 through E5. Use the BANK tab to select one bank or all banks; this will
automatically start a download operation. Alternatively, simply click on a column to
initiate a download operation. The MENU headings are described below:
FILE: Save the event data in a file
BANK: Select one bank or all banks to initiate a download from meter to PC
PRINT: Click to print the data to the local printer
EXIT: Close the EVENT window
The Parameters shown in the left-most column of the EVENT window are explained
below:
USED OR NOT: Indicates that there are or are not data records in a particular
bank. The word RECORDING will appear if the meter is currently performing a
noise survey.
Criterion, Threshold, Exchange, and Time weighting are the settings configured by
the user for a noise survey as described previously.
115VdB RMS: Alerts the user that the sound level reached 115dB
EXCEED 140dB: Alerts the user that the sound level reached 140dB
START DATE and TIME: Date and time when noise survey began
STOP TIME: Time of day noise survey ended
EXPOSURE TIME: Duration of test
DOSE VALUE: Dose value in percent for the period of the test
TWA: Projected sound level in dB. In the example above, the elapsed time of the
noise exposure test is only 17 minutes; much shorter than an 8-hour test. However,
if sound levels remain relatively constant for the next 8 hours, the projected sound
level average will be 57.2 dB.
PEAK FLAG TIME: Time of day when sound level reached 115dB
PEAK DURATION: Interval of time that sound level was equal to or greater than
115dB