EX230-EU-EN-V1.2-09/10 13
Function Range Resolution Accuracy
Resistance
400 Ω 0.1 Ω ±(0.8% reading + 5 digits)
4 kΩ 0.001 kΩ
±(0.8% reading + 2 digits)
40 kΩ 0.01 kΩ
400 kΩ 0.1 kΩ
4 MΩ 0.001 MΩ
±(2.5% reading + 8 digits)
40 MΩ 0.01 MΩ
Capacitance 40.00 nF 10 pF
±(5.0% reading + 7 digits)
400.0 nF 0.1 nF
4.000 µF 1 nF
40.00 µF 10 nF
100.0 µF 0.1 µF
Note: Auto ranging; Input protection 600Vdc and AC rms
Frequency 5.000 Hz 0.001Hz
±(1.0% reading + 3 digits)
50.00 Hz 0.01 Hz
500.0 Hz 0.1 Hz
5.000 kHz 1 Hz
50.00 kHz 10 Hz
500.0 kHz 100Hz
5.000 MHz 1kHz
±(1.2% reading + 4 digits)
10.00 MHz 10kH
Notes: Sensitivity: >0.5V rms (≤1MHz); >3V rms (>1MHZ);
Input protection: 600V dc or ac rms
Duty Cycle 0.1 to 99.9% 0.1%
±(1.2% reading + 2 digits)
Notes: Pulse width: >100 µS <100ms; Frequency: 5Hz to150kHz;
Sensitivity: <0.5V rms; Overload protect: 600V dc or ac rms
Temperature
-4 to + 1400°F 0.1°F ±(3.0% reading +9°F)
-20 to +760°C 0.1°C ±(3.0% reading + 5°C)
Sensor: Thermocouple type ‘K’; Overload protection: 600V dc or ac rms
Temperature IR
-20 to 0°C
0.1°C/°F
±4°C
0 to 93°C ±(2.0%rdg or 2°C)
93 to 230
±(3.0%rdg or 3°C)
-5 to 32°F ± 8°F
32 to 200°F
±(2.0%rdg or 4°F)
200 to 446°F
±(3.0%rdg or 6°F)