HD300-EU-EN V2.2 07/13
12
Range Specifications
Air Velocity Measurements Range Resolution Accuracy (%rdg)
m/s (meters per second) 0.40 – 30.00 m/s 0.01 m/s ± (3% + 0.20m/s)
km/h (kilometers per hour) 1.4 – 108.0 km/h 0.1 km/h ± (3% + 0.8km/h)
ft/min (feet per minute) 80 – 5900 ft/min 1 ft/min ± (3% + 40ft/min)
mph (miles per hour) 0.9 – 67.0 mph 0.1 mph ± (3% + 0.4m/h)
knots (nautical miles per hour) 0.8 to 58.0 knots 0.1 knots ± (3% + 0.4knots)
Air Flow Measurements Range Resolution Area
CMM (cubic meters per minute) 0-999,900 m
3
/min 0.001 0 to 999.9m
2
CFM (cubic feet per minute) 0-999,900 ft
3
/min 0.001 0 to 999.9ft
2
Temperature Range Resolution Accuracy
Air Temperature (vane) -10 to 60
o
C (14 to
140
o
F)
0.1
o
C/F ±2
o
C (4
o
F)
-50 to -20
o
C (-58 to -
4
o
F)
0.1
o
C/F ± 5
o
C (9
o
F) Surface Temperature (IR)
-20 to 500
o
C (-4 to
932
o
F)
0.1
o
C/F ±2% or ±2
o
C (4
o
F)